 Purpose: To compare Orbscan II and Pentacam keratometry readings in terms of their agreement with a manual Javal type keratometer

 Methods: In this retrospective study, records of patients who had refractive surgery were reviewed. We extracted data of 765 eyes which had keratometry with the Javal keratometer of these, 577 had Orbscan II and 200 eyes had Pentacam acquisitions. Minimum (min-K) and maximum (max-K) keratometry readings and keratometric astigmatism with the latter two devices were compared with Javal.

 Results: Correlation coefficients for Javal and Orbscan II in measuring min-K and max-K were r=0.916 and r=0.913, respectively (p<0.001). The 95% limits of agreement (LoA) between Javal and Orbscan II was 1.17-1.20 D for min-K and 1.22-1.24 D for max-K. The coefficients for Pentacam and Javal min-K and max-K readings were very high (r=0.943 and r=0.962). The 95% LoA between Pentacam and Javal in measuring min-K and max-K were 0.51-0.99D and 0.72-0.99D, respectively. The correlation between Pentacam and Javal measurements of keratometric astigmatism was stronger than that for Orbscan II and Javal (r=0.973 and r=0.800) the 95% LoA was 0.55-0.76D for Pentacam and Javal, and 1.14-1.19D for Orbscan II and Javal.

 Conclusion: According to this research, Orbscan II and Pentacam had high correlation and agreement with Javal-keratometer in determining keratometric values. Nevertheless, the results obtained from Pentacam showed better agreement and stronger correlation with Javal as compared with Orbscan II. It seems that Pentacam is a suitable substitute for Javal to perform keratometry in normal eyes.

  

 Iranian Journal of Ophthalmology 201426(1):3-10 © 2014 by the Iranian Society of Ophthalmology

Keywords: Keratometry, Orbscan II, Pentacam, Javal-keratometer
Full-Text [PDF 433 kb]   (2445 Downloads)    
Type of Study: Original | Subject: Anterior segment
Send email to the article author

Add your comments about this article
Your username or Email:

Write the security code in the box >



XML     Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Hashemi H, Yekta A, Ostadimoghaddam H, Norouzirad R, Khabazkhoob M. Comparison of Keratometric Values Using Javal Keratometer, Oculus Pentacam, and Orbscan II. Iranian Journal of Ophthalmology. 2014; 26 (1) :3-10
URL: http://irjo.org/article-1-900-en.html
Volume 26, Issue 1 (Iranian Journal of Ophthalmology 2014) Back to browse issues page
مجله چشم پزشکی ایران Iranian Journal of Ophthalmology
Persian site map - English site map - Created in 0.056 seconds with 794 queries by yektaweb 3571