The Accuracy of Ocular Sonography in Detection and Measurement of Intraocular Foreign Bodies
دكتر محسن فروردین، دكتر مرسل مهریار، دكتر محمد کران جم، دكتر حسین اشرف، دكتر مرتضی مهدیزاده، دكتر منصور رحیمی، دكتر حمید حسینی، دكتر سیروس سیوفی*، دكتر مهناز مصلائی
، siyoofisirus@yahoo.com
چکیده:   (11075 مشاهده)

Purpose : To evaluate the accuracy of ocular sonography in measurement of size of intraocular foreign bodies (IOFB)

Methods : In a case series study, 58 consecutive patients with intraocular foreign body were included. Ocular sonography was performed to detect intraocular foreign body and the results were compared with the same intraoperative findings.

Results : Metallic material was found to be the most prevalent one (71%) followed by glasses (12%), stone (8%) and wood (6%). Ultrasound localized the foreign body lodged in the lens in 10.3%, in the angle in 3.4% and in the vitreous or vitreoretinal region in the rest of them (86.3%). The mean size measured by ocular sonography was 3.5×4.5 mm and that was reported by the surgeon was 2.9×4 mm. While the reports of sonography in detection and localization of IOFB was completely comparable with surgical findings, size measurements were considerably different from true measurements after intraocular surgery.

Conclusion : Ocular sonography showed mostly overestimation in measuring intraocular foreign bodies. Although ocular sonography was one of the best methods for assessment of IOFB, it doesn't have high accuracy in estimation of size of IOFB.

 

Iranian Journal of Ophthalmology 200820(4):20-23
متن کامل [PDF 49 kb]   (2913 دریافت)    
نوع مطالعه: پژوهشی | موضوع مقاله: عمومى
ارسال پیام به نویسنده مسئول

ارسال نظر درباره این مقاله
نام کاربری یا پست الکترونیک شما:

CAPTCHA


XML   English Abstract   Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Farvardin M, Mehryar M, Karanjam M, Ashraf H, Mehdizadeh M, Rahimi M, et al . The Accuracy of Ocular Sonography in Detection and Measurement of Intraocular Foreign Bodies. Iranian Journal of Ophthalmology. 2008; 20 (4) :20-23
URL: http://irjo.org/article-1-187-fa.html

فروردین محسن، مهریار مرسل، کران جم محمد، اشرف حسین، مهدیزاده مرتضی، رحیمی منصور، و همکاران.. The Accuracy of Ocular Sonography in Detection and Measurement of Intraocular Foreign Bodies. مجله چشم پزشکی ایران. 1386; 20 (4) :20-23

URL: http://irjo.org/article-1-187-fa.html



جلد 20، شماره 4 - ( مجله چشم پزشکی ایران 1386 ) برگشت به فهرست نسخه ها
مجله چشم پزشکی ایران Iranian Journal of Ophthalmology
Persian site map - English site map - Created in 0.06 seconds with 31 queries by YEKTAWEB 4072